S-EVAP-KD
Analytical Solvent Evaporators for
Kuderna-Danish Flasks
![]() S-EVAP-KD Catalog No. 120115-C8 |
Features:
|
| The instrument consists of a base and stand assembly, mounted in a deep bath. The entire instrument rotates so that all samples are accessible from the front. Kuderna-Danish flasks sit on a cover disk with attached concentrator tubes immersed in the bath water. Solvents with boiling points up to 65° C can be handled. An optional vacuum manifold permits solvents withboiling points up to 90° C to be processed. | |
| No. of Sample Positions | 8 | 5 | 8 | 5 |
|---|---|---|---|---|
| Instrument Cat. No. | 12025 | 12035 | 12015 | 12045 |
| Bench space (w x d x h) |
24x22x40 in (61x56x102 cm) | 26x24x40 in (66x61x102 cm) | ||
| Solvent Flask Type | Kuderna-Danish | |||
| Largest Size | 250 ml | 500 ml | 500ml | 1000ml |
| Concentrator TubeTypes and Size | 10 ml
non-insulated 10 ml vacuum insulated |
|||
| Condenser Type and Size | Inverted
Hopkins 260 mm 24/40 24/40 |
|||
| Individual Collection Flask Type and Size | Round with
Flat Bottom 250 mi |
Round with
Flat Bottom 500 mi |
||
| Central Collection Flask Type and Size |
8 port 2500 ml |
5 port 2500 ml |
8 port 4000 ml |
5 port 4000 ml |
| Water Flow-Meter Range | O to 3.0 lpm | |||
| Inside Dia. & Depth | 12x8.5 in (31x22 cm) | 16x8.5 in (41x22 cm) |
|---|---|---|
| Outside Dia.& Height | 16x11 in (41x28 cm) | 20x11 in (51x28 cm) |
| Total Watts | 1100 W | 1200 W |
| Voltage | 120 or 240 V AC | |
| Hertz | 50 or 60 Hz | |
| Temp. Range | 30° to 100° C | |
| Accuracy | ±2°C | |
| The condenser water supply Includes a flow meter, rigid supply tube, rotating inlet/drain port, parallel supply/return manifolds and a flexible return tube. Quick-disconnect fittings are used on the manifolds. The system can be connected to a bench cold-water tap. A refrigerated water chiller is recommended as an accessory for improved control of condenser temperature and water conservation. |
| The 8.5-inch deep, stainless steel water bath is thermostatically controlled and will accurately maintain water temperature from 30° to 100° C. It is equipped with a water level device, a high temperature limit switch and a GFCI electrical plug for safety. An intrinsically safe, purged case is available as an option. The bath is UL/CUL approved for laboratory use. |
| Options: | Suffix: |
|---|---|
| · Central collection glassware and tubing | (-C5, -C8) |
| · Reduced height glassware | (-S) |
| · Vacuum manifold and tubing | (-V) |
| · Intrinsically safe, purged bath case | (-Z) |
| · 240 volt electrical wiring and plug | (-2) |
Please add the appropriate suffix when ordering